Medtronic Recalls HeartWare Splice Kit

Published On 2017-05-08 09:39 GMT   |   Update On 2017-05-08 09:39 GMT

Medtronic is recalling the driveline splice kit due to a design problem that would prevent the repaired cable assembly from withstanding excessive force or pull (e.g. accidental dropping of controllers or snagging driveline cables). The excessive force or pull could cause damage to the cable assembly and interrupt electrical connection. The company issued these kits between April 2010 and March 2015 to repair previous HVAD driveline cable connector assembly issues.


An interruption in electrical connection may cause the pump to stop, which may lead to serious adverse health consequences, including death.


The HeartWare Ventricular Assist Device (HVAD) helps deliver blood from the heart to the rest of the body. It is used in patients who are at risk of death from end-stage left ventricular heart failure and who are waiting for a heart transplant. The system includes a pump implanted in the space around the heart (pericardium). It also includes a driveline cable that goes through the skin. The driveline connects the pump to an external controller that regulates the speed and function of the pump. The system is powered with batteries or power adapters. The HVAD is designed for use both in and out of hospital settings, including during patient transport.


The driveline splice kit is intended to repair the driveline once an electrical break has been identified. These repairs are limited to the section of the driveline that is outside of the patient.


Those who may be affected: Healthcare providers who used the HeartWare Ventricular Assist System Pump Driveline Splice Kit prior to April 2015. All patients that underwent driveline splice repairs prior to April 2015 should.




  • Determine whether performing a re-splice procedure is appropriate for patients who had a driveline repair performed with the original splice kit.

  • Healthcare providers should weigh the benefits of using the new HVAD Driveline Splice Kit against the risks of a re-splice procedure, during which the pump will temporarily have to stop. This may expose the patient to the risk of serious adverse events or death.

  • If the decision is made to perform a driveline re-splice, healthcare providers should contact their local HeartWare representative to schedule the procedure. The driveline splice procedure must only be performed by a minimum of two trained HeartWare technicians in a controlled clinical setting to reduce the potential risk of damage to the driveline and subsequent serious injury or death.

  • All patients that underwent driveline splice repair after April 2015 have been spliced with the improved HVAD Splice Kit and are not affected by this issue.


Healthcare professionals and patients are encouraged to report adverse events or side effects related to the use of these products to the FDA's MedWatch Safety Information and Adverse Event Reporting Program.

Article Source : FDA

Disclaimer: This site is primarily intended for healthcare professionals. Any content/information on this website does not replace the advice of medical and/or health professionals and should not be construed as medical/diagnostic advice/endorsement or prescription. Use of this site is subject to our terms of use, privacy policy, advertisement policy. © 2020 Minerva Medical Treatment Pvt Ltd

Our comments section is governed by our Comments Policy . By posting comments at Medical Dialogues you automatically agree with our Comments Policy , Terms And Conditions and Privacy Policy .

Similar News